2. Click Import.
3. In the Open dialog box, browse for the file you want to import, then click Open. The
routine becomes the active routine.
Exporting a Routine
1. Open the Maintenance page, then open the Cmds & Routines tab.
2. Click Export.
3. In the Save As dialog box, navigate to the folder where you want to store the routine file,
then click Save.
NOTE: If you haven't previously saved a routine, the Export button will not
display.
Probe and Heater Tab
Maintenance > Probe & Heater
Use this tab to adjust the probe height and plate heater settings.
Auto Adjust Height - Automatically adjusts the probe height using the specified location(s)
in the main plate, reservoir, and strip wells.
NOTE: Luminex recommends that you calibrate the probe height using well
D6 for a 96-well plate and well H12 for a 384-well plate.
Maintenance Page
101